GE MEDICAL SYSTEMS
D
IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Chapter 4 - Functional Checks  4-19
4-4-3 C358 Curved Probe Image Quality Tests
4-4-3-1 2D Artifacts Test
4-4-3-2 CFM Artifacts Test
Table 4-30    2D Artifacts Test (C358)
Step Task Expected Result(s)
1
Use the Standard Imaging Phantom 
RMI403GS.
2
Set the following parameters:
• Depth: 22cm
• Frame Rate: 2 (single focus)
• Focus Location: 12cm
3
Look for artifacts in the image.
Table 4-31    CFM Artifacts Test (C358)
Step Task Expected Result(s)
1
Use the Electronic Flow Phantom RMI1425A.
Select the Abdomen preset from the 
Abdomen preset options, and set the 
2D Gain to 45.
2
Select CFM and set the following CFM 
parameters:
• Gain: Maximum possible without color 
noise.
• Frame Rate: 2
• LVR: Minimum setting
3
Place the C358 probe on the Phantom 
(top view A).
Ensure that there are no flash strikes or other color artifacts.