IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Chapter 3 - Installation 3-61
5) If the connection was successfully established, trackball to the Dicom Storage Server line and
press Select, then trackball to the Check button and press Select.
If the connection is established, a green checkmark is displayed.
If the connection cannot be established, a red cross is displayed.
Properties of the DICOM storage server configuration are displayed on the right side of the tab, as reference.
6) If the connection was successfully established, trackball to the Dicom Print line and press Select,
then trackball to the Check button and press Select.
If the connection is established, a green checkmark is displayed.
If the connection cannot be established, a red cross is displayed.
7) If all connections were established successfully, click OK to exit the configuration procedure. If not,
review the entire procedure to locate the error.