GE MEDICAL SYSTEMS
D
IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
5-42 Section 5-6 - External Peripherals
Section 5-7
Vivid™ 3 Power Distribution
5-7-1 Electrical Power
The Vivid™ 3 system can be set up to operate from the following 50/60 Hz AC voltages: 100V, 120V,
220V, 230V and 240V.
The local input voltage is converted to an internal voltage of 230V, via the Trafo. This voltage is
distributed to different system components and converted into DC voltages in the BEP and the Front
End, to support their operation.
Table 5-9 below shows which voltages are used in the various system
components, both in the BEP and the Front End.
Table 5-9 Vivid™ 3 - DC Voltage Distribution
End
F = Front
B = Back
Board / Unit
+3.3V
Dig
+5 V
Dig
+5V
Ana
+12V
TEE
-5 V
Vee
10 V
+15V
-15V
Pmx85 n
Pmx85 p
+80 V
+
40 V
5 V
BE
12 V
BE
Local
42V
LVPS
230V
AC Int
Local
AC
Voltage
3.3 V
BE
F Mother Board
+ + + + + + + +
F Probes
+ +
F F Board
+ + + + + +
F MUX Card
+ + + + + + +
F BF
+ + + +
F FEC
+ + +
F RFT
+ +
F IMP
+ +
F RFI
+ + +
B KB Controller
+
B KB Matrix
+ +
B VIC
+ +
F DC LVPS
+ + +
F TXPS
+
Peripherals
+
Display
+
F Fans
+
B
Analog
Keyboard
+ +
B BE
+ + + +
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com