IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Chapter 6 - Service Adjustments 6-13
Note: Always perform the calibration in similar lighting conditions to those where the unit will be
used. If the unit is to be used in different lighting conditions, perform full calibration in both
dark and light conditions. It is suggested that you write down the brightness and contrast
settings for both conditions, so that the monitor can be reset to the calibrated values.
3.) Press the Exit button twice to close the on-screen menu. The scanning screen is displayed.
4.) Perform monitor calibration as described in the
Monitor Calibration on page 6 - 16.