IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
7-16 Section 7-3 - Performing Front End (FE) Diagnostics
3) In the VME tab, verify that the Comment area indicates that the system is ready to perform the test.
4) Trackball to the Vme Com Test button and press Select. The Vme Com Test LED changes color
as follows:
Yellow: The test is in progress.
Green: The test has been completed and no problems were identified.
Red: Problems were identified during the performance of the test.
5) Review the test results in the Comment area and proceed as follows:
If no problems were identified, proceed to the RFT test, as described in the
RFT Diagnostic
Tests (for RFT Configuration)
section, on page 7-17.
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
Calibration on page 6 - 14, and then repeat the VME test.
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
Chapter 5 - Components
and Function (Theory)
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.