EasyManua.ls Logo

GE Vivid 3

GE Vivid 3
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
7-18 Section 7-3 - Performing Front End (FE) Diagnostics
3) In the RFT tab, verify that the Comment area indicates that the system is ready to perform the test.
4) Trackball to one of the following buttons and press Select:
Select the Board Test button to perform Self Test 1 and Self Test 2 sequentially.
Select the Self Test 1 button to perform Self Test 1 independently.
Select the Self Test 2 button to perform Self Test 2 independently.
The relevant LED changes color as follows:
Yellow: The test is in progress.
Green: The test has been completed and no problems were identified.
Red: Problems were identified during the performance of the test.
5) Review the test results in the Comment area and proceed as follows:
If no problems were identified, proceed to the FEC test, as described in the
Front End Controller
(FEC) Diagnostic Tests (for RFT Configuration)
section, on page 7-19.
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in the
Calibration section, on page 6-14, and then repeat the RFT test.
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
Chapter 5 - Components
and Function (Theory)
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for GE Vivid 3

Related product manuals