IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
7-24 Section 7-3 - Performing Front End (FE) Diagnostics
7) Proceed as follows:
• If no problems were identified, proceed to the FB tests, as described in the
Front Board
Assembly (FB) Diagnostic Tests
section, on page 7-25.
• If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
• If errors were identified but the results were marginal, perform calibration, as described in the
Calibration section, on page 6-14, and then repeat the test.
• If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in Chapter 5 - Components
and Function (Theory)
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
8) When finished, trackball to the Cancel button and press Select. The FE Diagnostics dialog box is
re-displayed.