GE MEDICAL SYSTEMS
D
IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
7-52 Section 7-3 - Performing Front End (FE) Diagnostics
Figure 7-29 Report Dialog Box - Normal Channels Test Low Frequency
The Report dialog box contains the following information:
• Test Result Message: Displays the test result.
• TR4 Modules: Displays a graphic representation of the front board (four groups of TR4
sockets). Each group consists of four TR4 sockets, each socket containing four vectors
(channels), as follows:
• Each vector is represented by a button (highlighted in green) that displays the number of
the socket to which the TR4 is connected.
• Each button is highlighted yellow when selected.
• Each button is highlighted red if errors were detected for one or more channels.
• Checkboxes: Enable you to manipulate bad channels:
• Show Only Bad Channels: Select this checkbox to display only bad channels in the
Display area.
• Mark bad channel: Select this checkbox to manually mark the currently selected channel
as bad.