GE MEDICAL SYSTEMS
D
IRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
2-10 Section 2-3 - Facility Needs
2-3-4-4 DICOM Option Pre-Installation Requirements
To configure the Vivid™ 3 ultrasound unit to work with other network connections, the network
administrator must provide the required information, which should include the following:
Vivid™ 3 Details: DICOM network details for the Vivid™ 3 unit, including the host name, local port,
IP address, AE title and net mask.
Routing Information: IP addresses for the default gateway and other routers in use at the site.
DICOM Application Information: Details of the DICOM devices in use at the site, including the
DICOM host name, AE title and IP addresses.