EasyManua.ls Logo

GE Voluson 730Expert - Section 3-6 Completing the Set Up

GE Voluson 730Expert
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-22 Section 3-6 - Completing the Set Up
Section 3-6
Completing the Set Up
3-6-1 Connecting the Unit to a Power Source
The connection of the Voluson® 730Expert ultrasound unit to a power source should be performed by
a qualified person who has completed basic Voluson® 730Expert System User Training.
Use only the power cords, cables and plugs provided by or designated by GE Healthcare - Kretztechnik
to connect the unit to the power source.
3-6-2 Power On / Boot Up
3-6-2-1 Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Connect the Main Power Cable to a hospital grade power receptacle with the proper rated voltage.
Never use an adapter that would defeat the safety ground.
3.) CPN6 only - Switch ON the Circuit Breaker (F1) and the Power Switch of peripherals (F2).
4.) CPN80-81 only - Switch ON the Circuit Breaker at the rear of the system.
CAUTION
!! CA UTIO N: Prio r to
Prior to connect the Voluson® 730Expert unit to a power source, verify compliance with all
electrical and safety requirements. Check the power cord to verify that it is intact and of
hospital-grade. Products equipped with a power source (wall outlet) plug should be connected
to the fixed power socket that has a protective grounding conductor. Never use an adapter or
converter to connect with a power source plug (e.g., a three-prong to two-prong converter).
WARNINGWARNING
!! WARNING:
The unit’s power must be supplied from a separate, properly rated outlet to avoid risk
of fire. See Section 2-2-2-1 "Voluson® 730Expert Power Requirements" on page 2-3 for
rating information. The power cord should not, under any circumstances, be altered to
a configuration rated less than that specified for the current.
CAUTION
!! CAUTION:
Whenever disconnecting the Voluson® 730Expert system from the electrical outlet, always
observe the safety precautions. First unplug the main power cable from the wall outlet socket,
then from the unit itself. Remove by pulling on the cable connector - DO NOT pull on the cable.
CAUTION
!! CAUTION:
The Voluson® 730Expert requires all covers!
Operate this system only when all board covers and frame panels are securely in place.
The covers are required for safe operation, good system performance and cooling purposes.
Figure 3-22 Circuit Breaker at rear of system with CPN6 installed
NOTICE
!! NOTICE:
When AC power is applied to the scanner, the ON/OFF switch on the Control panel is illuminated,
indicating that the System (including the Back-end Processor) is in standby mode.
Circuit Breaker
Connector for
Mains Power Cable
Power
Switch of
Peripherals

Related product manuals