General Ultrasound system requirements
Ultrasound System – Common Service Information 3-5
Direction 5444964-100 English Rev. 5
Electrical requirements
General requirements
NOTE:  GE requires a dedicated power and ground for the proper 
operation of its Ultrasound equipment. This dedicated power 
shall originate at the last distribution panel before the Ultrasound 
system.
The Ultrasound System will function on voltages from 100-240 
Volts and 50 or 60 Hz. However, if using 220 volt power in North 
America, then a center tapped power source is required.
Sites with a mains power system with defined Neutral and Live
:
The dedicated line shall consist of one phase, a neutral (not 
shared with any other circuit), and a full size ground wire from 
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral
:
The dedicated line shall consist of one phase (two lines), not 
shared with any other circuit, and a full size ground wire from the 
distribution panel to the Ultrasound outlet.
NOTE:  Please note that image artifacts can occur, if at any time within 
the facility, the ground from the main facility's incoming power 
source to the Ultrasound system is only a conduit.
Electrical requirements for the Ultrasound System
The electrical requirements vary depending on the mains 
voltage and the equipment in use. 
Contents:
•  ‘Electrical requirements for Console Ultrasound Systems’ on 
page 3-6
•  ‘Electrical requirements for Laptop Ultrasound Systems’ on 
page 3-11
•  ‘Electrical requirements for the Handheld Ultrasound 
Systems’ on page 3-15
•  ‘Electrical requirements for the Ultrasound workstations’ on 
page 3-16