Second step
Follow this step only if you are reloading all presets for a probe(s)
Go back into Utilities>system setup>Administration on the Lt menu
Service tab: password – she
Select Application settings – Mid right of the screen
Reselect the same file from CD/DVD or USB and select load – you will not be prompted to
select the different probes/presets
Back-up load is now complete