General Ultrasound system requirements
Ultrasound SystemCommon Service Information 3-5
Direction 5444964-100 English Rev. 5
Electrical requirements
General requirements
NOTE: GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
The Ultrasound System will function on voltages from 100-240
Volts and 50 or 60 Hz. However, if using 220 volt power in North
America, then a center tapped power source is required.
Sites with a mains power system with defined Neutral and Live
:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral
:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
NOTE: Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
Electrical requirements for the Ultrasound System
The electrical requirements vary depending on the mains
voltage and the equipment in use.
Contents:
‘Electrical requirements for Console Ultrasound Systems’ on
page 3-6
‘Electrical requirements for Laptop Ultrasound Systems’ on
page 3-11
‘Electrical requirements for the Handheld Ultrasound
Systems’ on page 3-15
‘Electrical requirements for the Ultrasound workstations’ on
page 3-16

Table of Contents

Other manuals for GE Voluson E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson E8 and is the answer not in the manual?

GE Voluson E8 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Power Requirements100-240 V AC, 50/60 Hz
ApplicationsObstetrics, Gynecology, Cardiology
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler
Transducer Ports4 active transducer ports
Image StorageInternal hard drive, USB storage
Transducer TypesLinear, Endocavitary
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals