Site Preparations
3-16 Ultrasound System – Common Service Information
Direction 5444964-100 English
 Rev. 5
Electrical requirements for the Ultrasound workstations
Contents:
• ‘Electrical Requirements - EchoPAC PC Turnkey 
Workstations’ on page 3-16
• ‘Electrical Requirements - LOGIQworks Workstations’ on 
page 3-16
Table 3-15:  Electrical Requirements - EchoPAC PC Turnkey Workstations
Product Voltage Tolerance
Current or 
Power 
Consumption Frequency
All models
100 VAC
±10%
8 A
50-60 Hz
110 VAC 8 A
230 VAC 4 A
Table 3-16:  Electrical Requirements - LOGIQworks Workstations
Product Voltage Tolerance
Current or 
Power 
Consumption Frequency
All models 110 - 230 VAC ±10% 310 to 543 W 
(depending on 
workstation 
model)