GenadyneBiotechnologies|EN(V5.0909.1) 18
TargetTimeout
The target timeout function is to enable the user to determine how soon after should the device start
alarming to notify the user of a leakage event occurring.
For example, if the target timeout is set at 30 seconds, this means that in the event of a target timeout,
it will have to last for at least 30 seconds continuously before the system starts alarming.
Use the Up/Down button to increase or decrease the desired time limit and press the Menu/Select
button to confirm setting.
AlarmLog
All alarms are logged and saved in the XLR8 memory.
To enter into the alarm log
1. Press Menu/Select button
2. Navigate to Alarm Setup by using the Up/Down button and press the Menu/Select button to
enter into the Alarm Setup function
3. Navigate to the Alarm Log by using the Up/Down button and press the Menu/Select button to
enter into the Alarm Log screen
4. All the past alarms will be shown on the screen
5. To acknowledged them scroll to the desired alarm notification and press the Menu/Select
button
6. The alarm bell will stop once acknowledged.
7. The asterisk (*) on the left side of the notification WILL NOT
disappear until the problem is
fixed.
To exit to the main screen, press and hold on to the Exit/Cancel button for 5 seconds.