Gendex Dental Systems
B-19
ORALIX AC SYSTEM INSTALLATION
3.4. INSTALLING THE FOLDING ARM
WARNING
The Folding Arm is packed in the folded
position and secured with tape; it is counter
loaded to balance the weight of the
Tubehead, therefore the utmost care must
be taken when removing the securing tape
(see also mounting Tubehead).
Lubricate the spigot (a) of the Folding Arm
with grease.
Loosen the four locking screws (b) a few
turns.
Push the ring (d) over the support pin (a),
fitting the ring hole into locking pin (e).
Pull down through the column the cable (c) of the Folding Arm to the mounting plate of the Timer.
Prepare the sector (f) on top of the column,
in the position indicated in the drawing.
WARNING
The installation of the sector is very important
for the proper stability of the Mobile Stand
and to ensure that the Folding Arm can be
extended only in the direction of the long legs
of the pedestal.
Insert the spigot of the Folding Arm into the
column of the Mobile Stand.
Ensure that the pin (e) is fitted into the sec-
tor hole (g).
Fasten the four locking screws (b).
(e)
(a)
(c)
(d)
(b)
l i g h t l y
grease
here
(e)
(f)
(g)
(b)
SET HERE
1
2
3
4
5
7
6