Before doing any precision levelling it is advised to
check the calibration of the instrument.
Set up the laser on a tripod at about 30m facing a
wall or staff with a detector.
Allow the instrument to level.
Detect and note laser position on the wall or staff.
Without moving the tripod rotate the instrument
180°.
Detect and note laser position on the wall or staff.
Calculate the distance between the two readings.
The difference should be within 6mm at 30m.
If the instrument is out of calibration it is advised
to send it in to Spot-on.
(see www.spoton.com.au for details)
CHECKING CALIBRATION