4 FA50 Operation Manual
Content
Requirements
inflammable gases, without heat source and
wind source, without mechanical vibration;
3) Good ventilation, avoid direct sunlight.
Space
Has enough space to place the analyzer,
with a minimum area of 0.8m
2
.
Environment
temperature
10-30
Relative
humidity
≤ 70%
Atmospheric
pressure
70.0kPa-106.0kPa
Ventilation
Able to exchange air with outside, smooth air
flow, no direct wind blowing to the analyzer.
Power supply
100-240V~, 50/60Hz, correctly grounded and
the grounding impedance is less than 0.1Ω.
Electromagnetic
wave
Do not close to the electric-brush engine or
electric devices which are frequently
switched on/off.
Ambient light
Avoid direct exposure to intensive light.