18
Instructions for use
5018610-00 Rev.D Edition 0703
Inspection
When necessary
Washing chamber
Keep the washing chamber free of limescale deposits. Remove any remaining
deposits as follows:
• Open the door and dispense 200 ml NEODISHER BU descaler (sold by Dr.
Weigert) into the wash chamber.
• Close the door and run the descaling program.
(NOTE: Do not have any goods in the machine with this program. It is only for descaling
the machine and insert.
The outside
The front and side walls can be cleaned and disinfected with an alcohol-based
cleaning agent, medical spirit or an ordinary cleaner. the control panel is cleaned in
the same way.
Remove stains with an ordinary cleaner for stainless steel.
If the washer disinfector is used without automatic descaling for incoming water,
the heating element must be checked regularly for limescale deposits.
It is important to maintain a high level of hygiene around the machine.
V1627