Important Servo-i Ventilator System
Revision 02 Service Manual 1 - 5
To the responsible service personnel
The contents of this document are not binding.
If any significant difference is found between the
product and this document, please contact
Maquet for further information.
We reserve the right to modify products without
amending this document or advising the user.
Only personnel trained and authorized
by Maquet shall be permitted to
perform installation, service or
maintenance of the Servo-i.
Only Maquet genuine spare parts must
be used. PC boards (spare parts) must always be
kept in a package for sensitive electronic
devices. Maquet will not otherwise assume
responsibility for the materials used, the work
performed or any possible consequences of
same.
The device complies to standards and
requirements as stated in the 'Servo-i Ventilator
System – User's manual'.
Installation
Only personnel trained and authorized
by Maquet shall be permitted to install
the Servo-i. The installation and
handing over procedures are described
in the 'Servo-i Ventilator System –
Installation Instructions'.
Functional check
After any installation, maintenance or service
intervention in the Servo-i, perform a 'Pre-use
check' according to instructions in the 'Servo-i
Ventilator System – User's manual'.
Service
The Servo-i must be serviced at regular
intervals by personnel trained and
authorized by Maquet.
Any maintenance or service must be
noted in a log book provided.
It is recommended that maintenance
and service is done as a part of a
service contract with Maquet.
Preventive maintenance must be performed at
least once every year as long as the unit is not
used more than normal. Normal operation is
estimated to correspond to approx. 5.000 hours
of operation. Details are found in this Service
Manual, chapter ”Preventive maintenance”.
The Battery modules shall be replaced every
three years.
The internal Lithium batteries (on PC 1771 and
PC 1772) shall be replaced every five years.
Worn-out batteries must be recycled or
disposed of properly according to local
regulations. Recycle facilities may not
be available in all areas.
Batteries must not be disposed of with
ordinary waste. Discard all other
disposable, replaced and left-over parts
in accordance with appropriate
industrial and environmental standards.
When working with ESD sensitive
components, always use a grounded
wrist band and a grounded work
surface. Adequate service tools must
always be used.
1
Important

Other manuals for Getinge Maquet SERVO-i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet SERVO-i and is the answer not in the manual?

Getinge Maquet SERVO-i Specifications

General IconGeneral
BrandGetinge
ModelMaquet SERVO-i
CategoryMedical Equipment
LanguageEnglish

Related product manuals