What to do if Getinge Maquet SERVO-i Medical Equipment panel button is stuck?
- JJennifer ScottAug 18, 2025
If the panel button on your Getinge Medical Equipment is stuck, try the following: * Restart the unit, avoiding any contact with the User Interface during startup. * Verify that System SW V1.03.00 (or later) is installed, and update if necessary. * Check the user interface membrane buttons (FSS). * Replace the touch screen. * Replace PC 1777 Panel.