User's Manual
SERVO-i/s
Cleaning and Maintenance

Other manuals for Getinge Maquet SERVO-i

Questions and Answers

  • R
    Rachel GonzalezAug 18, 2025
    What to do if Getinge Maquet SERVO-i Medical Equipment panel button is stuck?
    • J
      Jennifer ScottAug 18, 2025
      If the panel button on your Getinge Medical Equipment is stuck, try the following: * Restart the unit, avoiding any contact with the User Interface during startup. * Verify that System SW V1.03.00 (or later) is installed, and update if necessary. * Check the user interface membrane buttons (FSS). * Replace the touch screen. * Replace PC 1777 Panel.
  • T
    Todd BishopAug 20, 2025
    What to do if Getinge Maquet SERVO-i safe valve is open?
    • S
      Shelby CooleyAug 21, 2025
      If the safe valve is open on your Getinge Medical Equipment, you should: * Check inspiratory channel. * Replace safety valve pull magnet. * Replace PC 1784 Expiratory channel.
  • S
    Sarah MaldonadoAug 23, 2025
    What to do if Getinge Medical Equipment panel audio failed?
    • E
      Erin WeissAug 23, 2025
      If the panel audio on your Getinge Medical Equipment has failed, you can try: * Restarting the unit, making sure not to touch the User Interface during system start-up. * Checking that System SW V1.03.00 (or later) is installed and update System SW if required. * Replacing the loudspeaker. * Replacing PC 1777 Panel.
  • S
    Sarah AllenAug 26, 2025
    What to do if Getinge Maquet SERVO-i Medical Equipment shows INSP_VALVE_RANGE_ERR?
    • W
      wfoleyAug 26, 2025
      If your Getinge Medical Equipment displays the error INSP_VALVE_RANGE_ERR, you can try: * Replacing PC 1771 Control. * Replacing PC 1772 Monitoring. * Replacing the gas modules one at a time, checking after each replacement to see if the error code disappears.
  • P
    pturnerAug 27, 2025
    What to do if Getinge Maquet SERVO-i shows POWER_FAILURE_24_VOLTS_TOO_LOW?
    • J
      jamie30Aug 28, 2025
      If your Getinge Medical Equipment shows a POWER_FAILURE_24_VOLTS_TOO_LOW error, you should: * Replace PC 1778 DC/DC & Standard Connectors. * Replace the gas modules. Replace one gas module at a time and check that this technical error code will not appear.
  • J
    John AlexanderAug 30, 2025
    What to do if Getinge Maquet SERVO-i shows POWER_FAILURE_12_VOLTS_TOO_LOW?
    • M
      Michael WintersAug 30, 2025
      If your Getinge Medical Equipment displays a POWER_FAILURE_12_VOLTS_TOO_LOW error, you can try the following: * Check the status of the external battery (if connected). * Replace PC 1778 DC/DC & Standard Connectors. * Replace PC 1775 Plug-and-play back-plane.
  • G
    Gina Lopez DDSSep 1, 2025
    What to do if Getinge Medical Equipment shows POWER_FAILURE_MINUS_12_VOLTS_TOO_HIGH?
    • L
      Lisa OrtizSep 1, 2025
      If your Getinge Medical Equipment displays a POWER_FAILURE_MINUS_12_VOLTS_TOO_HIGH error, you can try: * Checking the status of the external battery (if connected). * Replacing PC 1778 DC/DC & Standard Connectors. * Replacing PC 1775 Plug-and-play back-plane.
  • S
    Sara FullerSep 4, 2025
    What to do if Getinge Maquet SERVO-i shows POWER_FAILURE_24_VOLTS_TOO_HIGH?
    • N
      Natalie LopezSep 4, 2025
      If your Getinge Medical Equipment shows a POWER_FAILURE_24_VOLTS_TOO_HIGH error, replace PC 1778 DC/DC & Standard Connectors.
  • J
    Jessica FreemanSep 6, 2025
    What to do if Getinge Maquet SERVO-i shows POWER_FAILURE_12_VOLTS_TOO_HIGH?
    • R
      Robert MejiaSep 6, 2025
      If your Getinge Medical Equipment shows a POWER_FAILURE_12_VOLTS_TOO_HIGH error, replace PC 1778 DC/DC & Standard Connectors.
  • H
    Hector PerrySep 8, 2025
    What to do if Getinge Maquet SERVO-i Medical Equipment shows POWER_FAILURE_MINUS_12_VOLTS_TOO_LOW?
    • W
      William ThomasSep 8, 2025
      If your Getinge Medical Equipment displays the error POWER_FAILURE_MINUS_12_VOLTS_TOO_LOW (i.e. < -13.2 V), replace PC 1778 DC/DC & Standard Connectors.

Summary

1 Introduction

1.1 Introduction

Provides an overview of the manual's scope and applicability to SERVO-i/s systems.

1.2 General Information

Discusses variations in cleaning practices and Maquet's recommendations.

1.3 Symbols

Explains the meaning of various symbols used throughout the manual.

1.4 Warning, Caution, Important and Note

Defines different types of information alerts to ensure safe operation.

2 Routine Cleaning

2.1 Cleaning;Disinfection Procedures - a summary

Offers a concise overview of cleaning and disinfection steps for the equipment.

2.2 Preparations and Dismantling

Details the steps for preparing and taking apart the equipment for cleaning.

2.3 Wiping and Discarding

Outlines procedures for wiping down equipment and properly disposing of parts.

2.4 Disinfection Procedures

Explains various methods for disinfecting the equipment components.

2.5 Sterilization Procedures (Not Recommended)

Discusses sterilization methods, noting they are generally not recommended.

2.6 Assembling

Provides instructions on how to reassemble the equipment after cleaning.

2.7 SERVO Ultra Nebulizer Cleaning

Details the cleaning process specific to the SERVO Ultra Nebulizer.

2.8 Aerogen Pro Nebulizer

Covers the cleaning procedures for the Aerogen Pro Nebulizer.

2.9 SERVO-i NAVA System Cleaning

Explains how to clean the SERVO-i NAVA system components.

2.10 SERVO CO2 Analyzer Cleaning

Provides instructions for cleaning the SERVO CO2 analyzer.

2.11 Accessories

Refers to cleaning routines for associated accessories.

3 Maintenance

3.1 Preventive Maintenance

Outlines recommendations for regular equipment inspection and maintenance.

3.2 O2 Sensor

Details maintenance requirements and notes for the O2 sensor.

3.3 O2 Cell

Covers the replacement procedure and safety warnings for the O2 cell.

3.4 Extended Cleaning of Inspiratory Channel

Provides instructions for advanced cleaning of the inspiratory channel.

3.5 Accessories

Refers to maintenance procedures for associated accessories.