2.7 SERVO Ultra Nebulizer cleaning
The Servo Ultra Nebulizer has been
discontinued.
The information in this document applies to
devices sold with versions 6.1 and earlier of
the SERVO-i Ventilator System.
Cleaning should be done after each patient or
according to hospital routine.
Perform a function test of the SERVO Ultra
Nebulizer after the cleaning. Refer to the
SERVO-i/s User's Manual.
Preparation
Wipe off/discard
Rinse
Disinfectant Autoclave
Washer-
Disinfector
Only the T-piece and nipples can be
autoclaved.
2.7.1 Preparation
• Disconnect the Servo Ultra Nebulizer from
the ventilator.
2.7.2 Wiping and discarding
• Unscrew the T-piece and discard the
medication cup.
The medication cup must be disposed of
in an environmentally safe way in
accordance with hospital regulations.
• Empty the buffer water from the nebulizing
chamber.
Important: Do not autoclave or use a
washer-disinfector when cleaning the
nebulizing chamber.