2.10 SERVO CO
2
analyzer cleaning
Cleaning should be done after each patient or
according to hospital routine.
Preparation
Rinse
Dry
Sterilization
Rinse
Disinfectant
Disinfectant
(wipe off)
Rinse
(wipe off)
2.10.1 Preparation
Disconnect the Capnostat sensor and
airway adapter from the ventilator.
2.10.2 Capnostat sensor
Disinfection procedure
Wipe the Capnostat sensor with a soft cloth
moistened in disinfectant (Cidex OPA or
isopropyl alcohol 70%).
24
SERVO-i/s, User's Manual
| 2
| Routine cleaning |