Sterilization with autoclave (NOT
recommended)
Dismantle
Wipe off/discard
Rinse
Dry
Dry
Autoclave
Assemble
Autoclaving will reduce the lifetime of the
expiratory cassette and is not recommend as
a cleaning method.
Sterilization is normally not necessary for the
expiratory cassette, as it is not an invasive
instrument, but when applied, use validated
processes only. See further instructions about
autoclaving on page 16.
2.2 Preparations and dismantling
D2D1
Dismantle
Wipe off/discard
Rinse
Disinfectant
Rinse
Washer-
Disinfector
Dry
Assemble
2.2.1 Preparations
Switch off the ventilator using the switch
on the rear of the User Interface.
Disconnect the ventilator from the power
and gas supply.
Disconnect any optional equipment from
the power supply and from the ventilator.
SERVO-i/s, User's Manual
9
| Routine cleaning |
2 |