2.7.6 Nebulizer membrane
Replacement of membrane if needed/used:
Peel off the old membrane.
Put a new membrane in place.
Make sure the new membrane is properly
attached. This is easiest to see from the
inside.
Check the SERVO Ultra Nebulizer for
leakage (Pre-use check, SERVO-i/s User's
Manual).
2.8 Aerogen Pro Nebulizer
Cleaning should be done after each patient or
according to hospital routine.
Perform a function test of the Aerogen Pro
Nebulizer after the cleaning. Refer to the
SERVO-i User's Manual.
Disinfectant Autoclave
Washer-
Disinfector
Rinse
Wipe off/discard
Preparation
Only the Aerogen Pro Nebulizer unit, T-piece
and neonate adapters can be autoclaved.
20
SERVO-i/s, User's Manual