2.1 Cleaning/disinfection procedures
- a summary
A summary of the procedures for cleaning,
disinfection and sterilization is given here.
Details for each step are given in the following
sections.
Cleaning when bacterial filter is used
Dismantle
Wipe off/discard
Assemble
Wipe off and discard filter, see page 11.
Disinfection when no bacterial filter is used
D2
D1
Dismantle
Wipe off/discard
Rinse
Disinfectant
Rinse
Washer-
Disinfector
Dry
Assemble
There are two disinfection procedures for the
expiratory cassette and associated parts, D1
and D2. In D1 (recommended) a
Washer-Disinfector (EN-ISO-15883-1) is used,
and in D2 a disinfectant is used.
8
SERVO-i/s, User's Manual
| 2
| Routine cleaning |