2.4 Disinfection procedures
D2D1
Dismantle
Wipe off/discard
Rinse
Disinfectant
Rinse
Washer-
Disinfector
Dry
Assemble
Important:
• The expiratory cassette is a precision
instrument and must be handled
carefully.
• After disinfection procedure, all parts
must be dried before use. If the
expiratory cassette is not dry it may not
pass the Pre-use check.
2.4.1 Washer-Disinfector (D1)
• Wash the parts with water only in a
washer-disinfector at a temperature of
85-95°C (185-203°F).
The following cleaning agent can be used:
• neodisher® MediClean forte
• Getinge MIS Detergent (for the expiratory
cassette)
1
• Place the expiratory cassette on its side
with the electrical connector (1) uppermost
as shown in the picture above.
Notes:
• The water pressure in the
washer-disinfector should not exceed
1.5 bar.
• The maximum water flow should not
exceed 10 l/min.
• The water must have free passage
through the cassette.
The expiratory cassette will last at least 200
cycles.
SERVO-i/s, User's Manual
13