2.6 Assembling
Rinse
Disinfectant
Rinse
Washer-
Disinfector
Dry
Assemble
Dismantle
Wipe off/discard
2.6.1 Re-inserting the expiratory cassette
SERVO-i
CLICK
SERVO-s
CLICK
• Hinge the expiratory cassette and press it
firmly down into lock position.
WARNING! Make sure that the
cassette clicks into position. Check that
it cannot be moved upwards and that the
button on top of the cassette is
completely ejected.
• Note on a log sheet that a routine cleaning
has been performed. Refer to hospital
guidelines.
2.6.2 PRE-USE CHECK
After cleaning, always perform a Pre-use
check. For more information, please refer to
the SERVO-i/s User's Manual.
If the Pre-use check fails:
• Carefully shake/tilt the cassette (5-7 times).
• Run the cassette in a ventilator system with
a test lung for 10 minutes.
• Re-do the Pre-use check.
SERVO-i/s, User's Manual
17