2.9 SERVO-i NAVA system cleaning
2.9.1 Wiping and discarding
Cleaning should be done after each patient or
according to hospital routine.
Wipe all parts with a soft lint-free cloth
moistened in soap and water or
detergent-based disinfectant.
Important: Do not immerse the Edi module
or the cable in fluid.
Note: In case of more contaminated
surfaces, use ethyl alcohol (70%) or
isopropyl alcohol (70%). Avoid contact with
electrical contacts.
Discard the Edi catheter. The Edi catheter is
for single patient use only.
Note: The Edi catheter must not be
disposed of with ordinary waste.