Wipe the nebulizing chamber and control
cable with a soft cloth moistened in soap &
water or detergent-based disinfectant.
Note: In case of more contaminated
surfaces, use ethyl alcohol (70%) or
isopropyl alcohol (70%).
2.7.3 Rinse
Rinse the parts thoroughly in water
(<35°C/95°F).
2.7.4 Disinfection procedure
Washer-Disinfector
Wash the T-piece and nipples in a
washer-disinfector at a temperature of
85-95°C (185-203°F).
Important: Do not put the nebulizing
chamber in a washer-disinfector.
Pre-disinfectant
Hexanios G+R
Aniosyme DD1
Gigazyme Plus
Important: Follow the disinfectant
manufacturer's recommendations and
instructions.
Disinfectant
Let the T-piece and nipples soak in a
disinfectant such as
- Cidex OPA
- Anioxyde 1000
Important: Follow the disinfectant
manufacturer's recommendations and
instructions.
Rinse the T-piece and nipples thoroughly
in water (<35°C/95°F).
2.7.5 Sterilization procedure
Autoclave
The T-piece and nipples can be autoclaved in
a validated process, typically at a temperature
of 134°C (273°F) for 4-7 minutes.
Check that the O-ring is correctly positioned.
Important: Do not autoclave the nebulizing
chamber.
SERVO-i/s, User's Manual
19