Troubleshooting Servo-i Ventilator System
6 - 12 Service Manual Revision 02
6
Error
code
Error message / Possible cause Recommended action
10 VALVES_DISABLED 1. Replace PC 1784 Expiratory channel.
2. Replace PC 1771 Control.
3. Replace PC 1778 DC/DC & Standard
Connectors.
4. Replace PC 1772 Monitoring.
11 SAFETY_VALVE_OPEN 1. Check inspiratory channel.
2. Replace safety valve pull magnet.
3. Replace PC 1784 Expiratory channel.
12 BRE_NODE_DISCON 1. Replace PC 1771 Control.
13 BRE_NODEID_CONFLICT N/A
14 PANEL_NODE_DISCON 1. Check the control cable
2. Replace PC 1777 Panel.
3. Replace PC 1778 DC/DC & Standard
Connectors.
Note: This error indicates communication failure
between PC 1777 and PC 1772 and the error
code will thus not be shown on the display (but
will be logged).
15 PANEL_NODEID_CONFLICT N/A
16 EXP_FLOW_MTR_NODE_DISCON 1. Replace PC 1784 Expiratory channel.
17 EXP_FLOW_MTR_NODEID_CONFLICT N/A
18 N/A N/A
19 N/A N/A
20 N/A N/A
21 N/A N/A
22 BUZZER_SILENCER 1. Replace PC 1772 Monitoring.
23 N/A N/A
24 BACKUP_CAP_ERR_MON 1. Replace PC 1772 Monitoring