Servo-i Ventilator System Troubleshooting
Revision 02 Service Manual 6 - 17
6
Error
code
Error message / Possible cause Recommended action
Expiratory Flow Meter
40001 Technical error in Exp flow meter.
If the technical alarm ‘Expiration flow meter
PC 1784 60 V underrange’ is logged together
with error code 40001, see recommended
action for this specific fault.
1. Replace the expiratory cassette.
2. Replace PC 1784 Expiratory channel.
3. Replace PC 1785 Expiratory channel
connector.
If technical alarm ‘Expiration flow meter PC 1784
60 V underrange’ is present:
1. Replace PC 1778 DC/DC & Standard
Connectors.
Expiratory cassette
Technical error in
Expiratory cassette
Technical error in Expiratory cassette or in the
communication with the cassette.
If the technical alarm ‘Expiration flow meter
Exp. cassette power failure’ is logged
together with error code ‘Technical error in
Expiratory cassette’, see recommended
action for this specific fault.
1. Replace the expiratory cassette.
2. Replace PC 1785 Expiratory channel
connector.
3. Replace PC 1784 Expiratory channel.
If technical alarm ‘Expiration flow meter Exp.
cassette power failure’ is present:
1. Replace PC 1778 DC/DC & Standard
Connectors.