2202 2304 Wasserdestillierapparate aus Glas
Glass Water Stills
Seite 12 von 28 Rev.02
Caution! Cooling water leaves
the still with a temperatureof up to 70°C.
Danger of scaldings!
4.5 Waste water outlet
When the function "Cleaning" (water exchange) is activated, water water from the boiler will
flow out through the connection waste water outlet. Connect a ½ " hose with a maximum
length of approx 1 metre. The hose for waster water outlet must by all means be led to a
drain on a lower level. It must have a slope on the complete length and must not be
obstructed. It must be secured that the hose can draw air at the end of the cleaning
procedure in order to cut off the water contact between two electrodes in the water still.
4.6 Drainage of the bi-stage boiler (only applicable for models 2302 and 2304)
Open the drain cock to withdraw mono distillate from the bi-stage boiler. During normal
operation, the drain cock must be closed. Only then will the bi-stage boiler be filled and
started.
Caution! Distilled water leaves
the still with a temperature
of more than 90°C.
Danger of scaldings!