97
u CO
2
ALM LO: adjust the lower limit of EtCO
2
alarm. If the measurement value is lower than CO
2
lower
alarm limit, the information “CO
2
TOO LOW” appears on the screen. After the measurement value returns to
the normal one, the information disappears.
u INS ALM HI: adjust the upper limit of InsCO
2
alarm. If the measurement value is higher than InsCO
2
upper
alarm limit, the information “INS TOO HIGH” appears on the screen. After the measurement value returns to
the normal one, the information disappears.
u AWRR ALM HI: adjust the upper limit of AwRR alarm. If the measurement value is higher than the upper
alarm limit of AwRR, the informationAWRR TOO HIGH” appears on the screen. After the measurement
value returns to the normal one, the information disappears.
u AWRR ALM LO: adjust the lower limit of AwRR alarm. If the measurement value is lower than the lower
alarm limit of AwRR, the informationAWRR TOO LOW” appears on the screen. After the measurement
value returns to the normal one, the information disappears.
u APNEA ALM: after selecting the alarm time for APNEA alarm (7 options: 10 s, 15 s, 20 s, 25 s, 30 s, 35 s,
and 40 s), the information “CO
2
APNEAwill appear on the screen after the corresponding time selected. The
alarm level is "HI".
OTHER SET: select it to enter "CO
2
SETUP" interface.
n WAVE SCALE: adjust the amplitude of CO
2
waveform display area, two options: “LO” or “HI”, the default value
is "LO".
n WORK MODE: change the work mode of CO
2
, two options: MEASUREMENT or STANDBY, the default is
"STANDBY". When you need to perform CO
2
monitoring, select "MEASUREMENT".
n ATMOS (mmHg): adjust current atmospheric pressure, range: 400 mmHg~850 mmHg, Resolution: 1 mmHg,
default:760 mmHg.
n O
2
COMPENSATE: set the gas compensation, which is used with "BALANCE GAS" and "ANEA" together.
Adjustable range: 0~100%, accuracy: 1%, default: 16%.
n BALANCE GAS: set the gas compensation, which is used with "O2 COMPENSATE" and "ANEA" together. Three
options: room air, N2O and Helium. The default is "room air".
n ANEA: set the gas compensation, which is used with "O2 COMPENSATE" and "BALANCE GAS" together.
Adjustable range: 0.0~20.0 %, accuracy: 0.1%, default: 0.0 %.
NOTE:
l Anesthetic gas will ignore it when the balance gas is set to "Helium".
n Zero: “Sample Cell Zero “is a quick process that allows the module to accommodate the optical characteristics of
the different adapter types. "Sample Cell Zero" should be performed whenever the type of adapter being used with
the module is changed. For optimal accuracy, "Sample Cell Zero "should also be performed whenever the module is
connected to the host system.
n DEFAULT: select it to enter "CO
2
DEFAULT CONFIG" menu, in which the user may select "FACTORY
DEFAULT CONFIG" or "USER DEFAULT CONFIG". After selecting one item and exiting the menu, the system
will pop up the dialog box asking for your confirmation.