129
A.13 TEMP
Measurement method
Thermistor method
Channel
Dual-channel
Probe type
YSI-2.252 K
Measurement site
Body surface probe: armpit
Body cavity probe: oral, rectum
Measurement range
0~50
Resolution
0.1
Accuracy
±0.1
Updating cycle
About 1 s
Response mean time
<10 s
Alarm response time
2 min
Unit
or
Alarm limit
Range(
)
Step(
)
T1/T2 high limit
(low limit +1)~50
1
T1/T2 low limit
0~(high limit-1)
TD high limit
0~50
A.14 IBP
Measurement method
Invasive and direct measurement
Channel
Dual-channel
Measurement range
-10~300 mmHg
Resolution
1 mmHg
Accuracy
±2 % or 1 mmHg, whichever is greater.
Update time
About 1 s
Pressure sensor
Sensitivity
5 uV/V/mmHg
Impedance range
300~3000 Ω
Unit
mmHg, kPa, cmH2O
Alarm limit
Range(mmHg)
Step(mmHg)
SYS ALM HI
(low limit+1)~350
1
MEAN ALM HI
DIA ALM HI
SYS ALM LO
-50~(high limit-1)
MEAN ALM LO
DIA ALM LO