T1 sensor is not connected
correctly.
Check the connection of T1 sensor.
T2 sensor is not connected
correctly.
Check the connection of T2 sensor.
larger interference signals appear
in the ECG signals.
Check the connection of ECG lead
cable; Check the current situation of the
patient. Check if the patient moves a lot.
XX module can't communicate
normally with the host.
Re-start up the monitor, if the error still
exists, contact the manufacturer.
XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, etc.
The keyboard has failures, which
cannot be used.
Contact the manufacturer for repair.
The network part in the system
appear problems, so the system
can't achieve network function.
Contact the manufacturer for repair.
"CAN'T FIND
WIRELESS
DEVICE"
Charge the battery by connecting the AC
cord.
Install the recording paper.
The recorder communicates
abnormally.
Turn off the device and restart it.
NIBP initialization error
Execute the reset program in the NIBP
menu. If the failure still exists, contact
the manufacturer for repair.
During NIBP measurement, illegal
reset occurs.
Check the airway of NIBP to check if
there are clogs. Then measure again, if
the failure still exists, contact the
manufacturer for repair.