specified upper safety limit.
measuring function of NIBP module and
notify biomedical engineer or Our service
staff.
Stop the patient from moving.
During pneumatic test, leak
is detected.
Check and replace the leaking parts, if
required, notify biomedical engineer or Our
service staff.
Operation of blood pressure
pump system is failed.
Stop using measuring function of NIBP
module, notify biomedical engineer or Our
service staff.
Cuff type does not comply
with the patient type.
Select an appropriate cuff type
Measurement time has
exceeded 120 seconds
(adult) or 90 seconds
(neonatal).
Measure again or use other measurement
methods.
The system cannot perform
measurement, analysis or
calculation during
measuring.
Check the cuff. Make sure that the patient
under monitoring is motionless. Measure
again.
Prompt message: (display in the prompt area below NIBP value)
During manual measuring mode.
During continuous measuring mode.
During automatic measuring mode.
After selecting interval time in MENU
Press NIBP key during measuring to stop measuring.