Chapter14 Troubleshooting
Abnormal
Phenomenons
Causes
Solutions
BP measurement
values too high or
too low.
Cuff is not connected correctly.
Correctly connect cuff (refer to
Chapter 9)
Move limbs when measuring
Keep quiet and restart a
measurement
No pressure
Cuff leakage
Buy a new cuff
The cuff windpipe is not
correctly connected with cuff
Correctly connect
Cuff is not inflated
Stop using the device and contact
us
Cuff deflates in
short time
Loose cuff
Correctly apply cuff
It can not carry on measurement when press the
measurement button
Switch on the power once again
and restart a measurement
Power off
suddenly when
inflating
No use for a long time, the
power of batteries can be
exhausted owing to the changed
Replace all four batteries with new
ones.