22
C. Operating current: <=40mA
D. SpO2 Parameter Specications:
Measuring range: 35% - 100%
Accuracy: ≤3% (during 70% - 100%)
E. Pulse Rate Parameter Specications:
Measuring range: 30bpm-240bpm
Accuracy: ±2bpm or ±2% (which ever is greater)
F. Blood Perfusion Parameter Specications:
Measuring range: 0%~20%
G. The performance under low perfusion condition
The accuracy of
SpO2
and PR measurement still meets the speci-
cation described above when the modulation amplitude is as low as
0.6%.
H. Resistance to ambient light interference:
The accuracy of
SpO2
and PR measurement still meets the speci-
cation described above when the device is tested by
SpO2
simulator
(Fluke Biomedical Index 2 series) while setting the emulating interfer-
ence of sun light and 50Hz/60Hz uorescent light.
I. Dimensions: 56 mm (L) x 47 mm (W) x 32 mm (H)
Net Weight: 45g (including batteries)
4. ACCESSORI
A. A hanging cord B. Two batteries
C. A charger D. A power adapter
E. A charger cable F. A pouch
G. A User Manual H. Quality Certicate