14
ENGLISH
RF emissions CISPR 11 Class B The Vein Finder is suitable for use in all
establishments, including domestic establish-
ments and those directly connected to the
public low- voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage uctuations/ icker
emissions IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration Electromagnetic Immunity
IEC 60601-1-2
The Vein Finder is suitable for use in the specied electromagnetic environment. The customer and/
or the user of the Vein Finder should assure that it is used in an electromagnetic environment as
described below:
Immunity Test IEC 60601-1-2
Test Level
Compliance Level Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete,
or ceramic tile. If oors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst IEC
61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output
lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be
that of a typical commercial and/
or hospital environment
Surge IEC 61000-4-5 ± 1 kV
differential mode
± 2 kV
common mode
± 1 kV
differential mode
± 2 kV
common mode
Mains power quality should be
that of a typical commercial and/
or hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% UT
(> 95 % dip UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
< 5 % UT
(> 95 % dip in
UT) for 5 sec
< 5 % UT
(> 95 % dip in
UT) for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
< 5 % UT
(> 95 % dip in
UT) for 5 sec
Mains power quality should be
that of a typical commercial and/
or hospital environment. If the
user of the Vein Finder requires
continued operation during
power mains interruptions, it
is recommended that the Vein
Finder be powered from an
uninterruptible power supply or a
battery.