16
ENGLISH
Recommended Separation Distances
between Portable and Mobile RF Communications Equipment and the Vein Finder
IEC 60601-1-2
Frequency of Trans-
mitter
Equation Rated
Maximum Output
Power
of Transmitter watts
150 kHz to 80 MHz
Separation Distance
metres
150 kHz to 800 MHz
Separation Distance
metres
800 MHz to 2,5
GHz
Separation Distance
metres
0.01
0,12
0,12 0,23
0.1 0,38 0,38 0,73
1 1,2
1,2 2,3
10 3,8 3,8 7, 3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the separation distance can be
estimated using the equation in the corresponding column, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reection from structures, objects, and People.