To disinfect, use any of the chemicals listed in Table 2 and do the
following:
1. Ensure the equipment is clean (see Cleaning Section for
Cobalt video baton).
2. Ensure the protective caps on the connectors are secure.
3. Prepare the disinfectant solution at the concentration and
temperature recommended by the disinfectant manufacturer.
Disinfect the equipment following the disinfectant
manufacturer’s instructions or as stated in Table 2.
4. After the disinfection process, rinse (as applicable) and dry
the Cobalt video baton and then store the equipment in a
clean environment.
Cleaning the Video Cables
Wipe the video cables with IPA (70% isopropyl alcohol), bleach
(100ppm) or a mild detergent and water.
High Level Disinfection, as stated in Table 2 is required for the
video cable when it is visibly soiled.
Cleaning the Monitor
Wipe the exterior of the monitor with IPA (70% isopropyl alcohol),
bleach (100ppm) or a mild detergent and water.
Cleaning the Cradle
Wipe the cradle with a standard hospital grade surface cleaning product.