What does Error E1 mean on GOK SmartBox 1 Measuring Instruments?
- LleahhooverSep 7, 2025
The error 'Error E1' on GOK Measuring Instruments indicates that the set value is invalid.
What does Error E1 mean on GOK SmartBox 1 Measuring Instruments?
The error 'Error E1' on GOK Measuring Instruments indicates that the set value is invalid.
What to do if GOK Measuring Instruments show Error E6?
If your GOK Measuring Instruments display Error E6, it means the current measured value is too low to be used as a reference point. Ensure the probe is submerged and that the set height isn't too high, which could cause a low measured value. Also, check menu item "9. Offset probe". If the issue persists, it may indicate a probe fault.
What to do if GOK Measuring Instruments show Error E3?
If you are getting Error E3 on your GOK Measuring Instruments, the measured value is too great for zero point calibration. Make sure the probe is not immersed during calibration.
What to do if GOK Measuring Instruments shows Error E5?
Error E5 indicates that the set height exceeds the tank's actual height. This is due to an incorrect entry in menu step 10. Please verify and correct the height setting in the menu.
What to do if GOK Measuring Instruments displays error E6?
Error E6 means that the current measured value is too low as a reference point. Ensure the probe is submerged and that the set height (menu step 10) isn't too high, causing a low measured value. Check menu item "9. Offset probe". Otherwise, the probe may be faulty.
What to do if GOK Measuring Instruments displays Error E4?
Error E4 means the measured value is not plausible. In this case, check menu item “9.Offset probe”.