PROGRAMMING
Press [Enter] to return to display mode
probe
Select probe measuring range (see type label of the
probe) - default value is 250 mbar
________mbar
Is only indicated if item 10. Set h
or 11. Set V has been activated
_______kg/m³
Measuring range kg/m³ (15 °C)
Measuring range kg/m³ (15 °C)
845 kg/m³ - default setting
Entry of a specific probe with
different measuring range
If the density of the stored medium is unknown, the reference height can be entered in menu
item “10.Trim height” To do this, determine the current fill level in mm, deduct 1.0 cm from it
and enter the resulting value - click YES to confirm the entry.
Should the current fill level be less than 75%, we urgently recommend that you adjust the
value after the next refuelling to achieve good measuring accuracy.