Gossen MetraWatt PROFITEST 204 - Symbols Which Appear at the Display of the PROFITEST 204 HP;HV High-Voltage Module

Gossen MetraWatt PROFITEST 204
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GMC-I Messtechnik GmbH 11
Data management message:
data transmission successfully completed
Excessive temperature inside the test instrument
(protective conductor test)
Allow instrument to cool down for approx. 10 minutes.
Internal temperature monitoring is defective.
Send the high-voltage module to the Repair Department
at GMC-I Service GmbH.
This symbol appears after the measured voltage
value has been frozen at the display (see chapter
4.4, page 33).
2.1.4 Symbols which Appear at the Display of the
PROFITEST 204HP/HV High-Voltage Module
High-voltage module ready for activation
A life endangering high-voltage of up to
5kV is present at the test probes of
the high-voltage module.
Test completed successfully:
residual voltage less than 25 V
The predefined limit value I
MAX
has been
exceeded. The current limiting function has
switched the instrument to the stand-by mode.
It is entirely possible that the displayed value for I
T
is less than I
MAX
, because the last measured value
is always displayed.
Instrument is defective! / Transmission
between PROFITEST 204+ and 204HP/HV
interrupted.
Excessive temperature inside the test instrument
(high-voltage test)
Allow instrument to cool down for approx. 10 minutes.