4.2.1 Configuring Test Parameters
Select a test voltage of 500 V for the main circuit, or 1000 V if
required for extended testing. A lower test voltage can also be
selected for testing voltage sensitive components. The allowable
limit value for insulation resistance can also be changed.
MENU
Select the desired parameter.
Adjust to the desired value.
4.2.2 Start the Test
START
or
Note
The test can only be started with the START key on the
test instrument from the test parameters window.
STORE
Press briefly: Store an event.
Press and hold:
Store an event and activate the entry field.
Up to 15 characters can be entered to the entry
field, or to the info field at the bottom left as a
description for the measurement.
See chapter 5.1.1 regarding data entry.
or
STORE
Press briefly: Store an event.
The green/red signal lamp is extinguished briefly as
acknowledgement.