4.5.4 Pulse Control Mode
The pulse control mode is recommended for troubleshooting
(locating the point of flashover).
Breaking current I
MAX
is set to a fixed value of approximately
125 mA for the 204HP and 5 mA for the 204HV in the pulse
control mode. If this value is exceeded, test voltage is switched
off within 0.5 ms. After approximately 0.6 s, test voltage is
increased from 0 to the selected maximum value within a period
of 0.2 s, or is switched off again if breaking current is exceeded.
Set I
MAX
to >>>>>>> for selection of the pulse control mode
(see chapter 4.5.2, page 37).
If continuous operation is required, set test duration to >>>>>>>
as well (see chapter 4.5.2, page 37).
4.5.5 Ending the Voltage Test
Release the triggers at the high-voltage pistols.
Press the MENU key.
The signal lamps switch from red to green.
If the test system is left unattended, it must be switched to the
out of service” mode (signal lamps off).
Turn the key switch to the “0” positions and remove the key.
Secure the instrument against unauthorized use.