Sensor oset is used in cases where there are deviations in the
cabinet’s actual operation compared to control measurements by
independent temperature measuring systems.
The A-sensor is used to manage the cabinets refrigeration system, and
is also the reference sensor for the display and alarms.
The A-sensor is oset if the actual temperature in the cabinet does not
match the setpoint, despite taking the hysteresis into consideration.
Oset of A sensor is named “CA”.
Osetting the A-sensor
Press and hold for more than 5 seconds
Press to select “CA”
Press + to oset the A-sensor
Press to conrm the set value
The A-sensor is now oset, proceed to other parameters by pressing
, and then navigate by using or
Leave the user menu by pressing several times until
the cabinet temperature is shown in the display.
Sensor oset