5. Specifications
The unit is designed for operation in cold rooms, incubators and closed laboratory rooms at ambient
temperature from +4ºC to +40ºC in a non-condensing atmosphere and maximum relative humidity 80% for
temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C.
5.1. Temperature specifications
Setting range ............................................................................................................+25°C ... +60°C
Control range .........................................................................................5°C above room t° to +60°C
Setting resolution ......................................................................................................................0,1°C
Stability....................................................................................................................................±0.1°C
Accuracy..................................................................................................................................±0.5°C
Uniformity at +37ºC ...............................................................................................................±0.25°C
Time of block heating from +25°C to +37ºC ...........................................................................12 min
5.2. General specifications
Speed setting range ...................................................................................................250–1200 rpm
Speed setting resolution .........................................................................................................10 rpm
Orbit...........................................................................................................................................2 mm
Display .....................................................................................................................16x2 signs, LCD
Digital time setting range ............................................................................1 min – 96 hrs / non-stop
Max. continuous operation time ..........................................................................................96 hours
(recommended interval between operation sessions not less than 8 hours)
Time setting resolution...............................................................................................................1 min
Max. height of microtest plate ................................................................................................18 mm
External power supply ............................................. input AC 100–240 V 50/60 Hz, output DC 12 V
PHMP PHMP-4
Current/power consumption 12 V DC, 3,3 A / 40 W 12 V DC, 4,15 A / 50 W
Number of microplates 2 4
Platform dimensions 250 x 150 mm 210 x 290 mm
Dimensions 270x260x125 mm 380x390x140 mm
Weight* 6.1 kg 8.8 kg
* Accurate within ±10%.
Grant is committed to a continuous programme of improvement and reserves the right to alter design
and specifications of the equipment without additional notice.
Page 9
Thermo shaker PHMP/PHMP-4
Operating instructions
Version 3.01 - May 2015