EasyManua.ls Logo

GRAPHTEC gl200a - 4.3 Performing Setup1 (Voltage Input)

GRAPHTEC gl200a
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GL200-UM-251-9370 4-5
4. SETUP PROCEDURES
4.3 Performing Setup1 (Voltage Input)
This procedure is required when any part of the AMP board is replaced.
Preparation
• Have a voltage generator on hand that can output voltage up to 50 VDC (four-digit output).
• Mutually connect the ground terminals of the voltage generator and the GL200/A.
• Set the AC Line cycle setting of the GL200/A to the same cycle as that of your AC line.
Wiring connection
0.00000 V
Voltage generator
FG
GND
AC adapter
Connect to CH1
Input cable
Connect both ends to GND

Related product manuals